Olympus B2-BDS-3 Troubleshooting & Adjustment Manual - Page 39
Browse online or download pdf Troubleshooting & Adjustment Manual for Microscope Olympus B2-BDS-3. Olympus B2-BDS-3 46 pages. For electrical base
D.
TROUBLESHOOT-ING '.'(B2..:..BDS-3)
·
No. D-8
4.
Power Supply
4-1
Power is not supplied even after the main switch is turned on,
SIOI
U I 0
Causes
1.
The breaker c:;BlOl functions.
2.
Defective main switch SlOl
3.
Defective AC cord
:
(UYCP15 in lOOV channel)
4.
Defective AC inlet PlOl
5.
Defective breaker CB101
6.
Defective filter PCB Ul04
7.
Defective primary PCB UlOl
Is the breaker functioning?
NO
0 -
--
------,
YES
Press the breaker button.
Is power supplied normally?
; !
YES
NO
.
'