Olympus BH-2 BHTU Complete Teardown, Cleaning, And Reassembly - Page 9

Browse online or download pdf Complete Teardown, Cleaning, And Reassembly for Microscope Olympus BH-2 BHTU. Olympus BH-2 BHTU 23 pages. Sliding focus block
Also for Olympus BH-2 BHTU: Manual (24 pages)

Olympus BH-2 BHTU Complete Teardown, Cleaning, And Reassembly
With  the  electrical  base  removed,  fragile  lighting 
components  are  exposed  on  the  open  bottom  of  the 
base which may be damaged if the base is not placed on 
a  suitable  clutter‐free  surface.    Place  the  base  in  its 
normal  upright  position  on  a  clean,  flat  surface 
someplace  out  of  harm's  way  and  where  dust  and 
debris  will  not  accumulate  on  the  exposed  lighting 
components.   
Disassemble the Sliding Focus Block 
Disassemble the sliding focus block mechanism per the 
following procedure. 
Remove the Trim Cover 
Use a suitable JIS screwdriver to remove the four M2x4 
screws  securing  the  U‐shaped  trim  cover  onto  the  arm 
casting (see Figure 8). 
Figure 8 – Remove screws securing the trim cover 
Remove  the  (now  loose)  trim  cover  from  the  arm 
casting (see Figure 9). 
Figure 9 – Remove the loose trim cover from the arm 
Loosen the Screws Tensioning the Slide Mechanism 
Use  a  2.5mm  Allen  wrench  or  driver  to  loosen  (but  do 
not  remove)  the  three  M3x12  hex,  socket‐head  cap 
screws  securing  the  guide  block  into  position  on  the 
back side of the focus block (see Figure 10). 
Complete Teardown, Cleaning, and Reassembly of the Olympus BH‐2 Sliding Focus Block  
 
Figure 10 – Loosen screws securing guide block in place 
Use  a  2.5mm  Allen  wrench  or  driver  to  loosen  (but  do 
not  remove)  the  two  M3x16  hex,  socket‐head  cap 
screws tensioning the slide mechanism (see Figure 11). 
 
Figure 11 – Loosen screws tensioning slide mechanism 
Remove the Focus Block Assembly from the Arm 
Carefully  slide  the  focus  block  (with  attached  guide 
block) out the bottom of the arm (see Figure 12), being 
careful to not lose any of the cylindrical rollers or spacer 
rods  in  the  process  (there  are  eight  rollers  and  one 
spacer rod on each side of the slide mechanism). 
 
Figure 12 – Withdraw focus block assembly from the arm 
Revision 2 