Explanations in detail
@
Mounting the halogen bulb
1 ) Releasing the bulb clamping levers
@
of
the collector lens
a
in the direction of
the arrow, insert two contact pins of the
halogen bulb into the socket
@.
(Fig. 1)
(Recommended t o use a glove or gauze
t o handle the halogen bulb.)
2 )
Secure the bulb in position with the two
levers.
*
Before use, wipe off any fingerprints or
stains on the bulb.
0
Mounting the stage
1) Prior to mounting the stage, rack down
the stage mounting dovetail
a
all the
way. (Fig.
2)
2)
Loosen the stage clamping screw
@
by
rotating counterclockwise
with
Allen
wrench provided. (Fig. 2)
Insert the stage into the mounting dove-
tail
a
from above slowly. (Fig.
2 )
3)
Lower the stage until it comes in contact
with the stop pin
0;
then clamp with
screw
@
.
(Fig.
2 )
0
Mounting the revolving nosepiece
Fig. 1
Fig.
2
1) Loosen the nosepiece clamping screw
a
(Fig.
3)
2 )
Aligning the nosepiece dovetall slide t o
the mounting block
@,
push in the nose-
piece
slowly all the way
*
Do not tilt or lock the nosepiece while
inserting into the mounting block.
Mounting the intermediate polarizing attach-
ment
1 ) Loosen the clamping screw
a
fully. Pull
spring-loaded clamping screw
@.
This
will cause the locating pin
@
to with-
draw. (Fig.
4)
If the pin does not, loosen
the screw further until the pin withdraws.
!
2 )
With clamping.screw
0
pulled out, insert
the circular dovetail of the intermediate
attachment into the ring dovetail.
3)
Tighten the clamping screw.
Fig.
3
Fig. 4