When you use BX3-SSU for the first time, be sure to perform the Compensation of the stage. For detail procedures, refer
to "4 Compensation of the Stage" described in the next page.
1
3
2
1
Downloaded from
ManualsNet.com
OPERATION OF SCANNING STAGE WITH ULTRASONIC
2
Fig. 2
2
1
Fig. 3
Fig. 4
search engine
1
Turning the Power ON
} Ensure that the cables are connected correctly and securely.
Set the main switches to " I " (ON) in the following sequence.
1. Set the main switch @ of the BX3-CBH/IX3-CBH control box to " I " (ON).
2. Set the main switch 2 of the control box for BX3-SSU to " I " (ON).
2
Moving the Stage
1. Rotate the Y-axis knob 2 and X-axis knob @ to move the stage.
Factory setting: 10 mm displacement per turn for the coarse adjustment,
0.2 mm displacement per turn for the fine adjustment.
2. Press the coarse/fine selector button 3 to switch the knob functions
between the coarse and fine adjustment.
· Pushed in: Coarse adjustment.
· Out position: Fine adjustment.
} The tone occurred when operating the stage may vary drastically in some
cases, but it does not cause any problem for the product performance.
} The stage may be vibrated slightly in some cases when starting the stage,
but the product is not damaged.
The stage or the slide holder may be interfered by the objective
CAUTION
lens depending on the type of the objective lens or the focus
position.
3
Replacing the Fuses
1. Set the main switch of the control box for BX3-SSU to "
unplug the power cord connector from the power outlet.
2. Draw out the fuse box @ by inserting a flat object such as a tip of
screwdriver into the slit 2 on the top surface of the fuse box.
3. Replace both of the fuses in the fuse box with new ones.
<< Applicable fuses >>
T3.15A(H)250V (LITTELFUSE 2153.15)
Using a non-specified fuse may cause a fire hazard.
CAUTION
4. Place the fuse box in the original position.
BX3-SSU
(Fig. 2)
(Fig. 3)
(Fig. 4)
" (OFF) and