Safety Symbols
The following symbols are found on the microscope. Study the meaning of the symbols and always use the equipment
in the safest possible manner.
Symbol
Caution indications
Caution indications are placed at parts where special precaution is required when handling and using the System. Always
heed the warnings.
Caution indication
position
1
Getting Ready
1. This manual pertains only to the reflected fluorescence system. Before using this system together with the BX3 microscope
and associated options, make sure that you have carefully read and understood their manuals, and understand how the
system should be operated together.
2. The reflected fluorescence system is composed of precision instruments. Handle it with care and avoid subjecting it to
sudden or severe impact.
3. Do not use the system where it is subjected to direct sunlight, high temperature and humidity, dust or vibrations.
4. To allow heat from the unit to dissipate well, reserve clearances of at least 10 cm around the lamp housing and power
supply unit.
5. The power cord can also be used to cut the power supply in case of emergency. To make this possible, the power supply
unit should be installed so that the power cord connector (on the rear of the power supply unit) or the power outlet is
easily accessible for unplugging in case of emergency.
6. Before igniting the burner, make sure that a mercury burner is attached and the cords are connected properly.
7. Avoid turning the lamp on-off repeatedly as this reduces the burner life considerably. The average service life of a mercury
burner is about 300 hours (USH103OL) with operating cycles of 2-hour ON and 30-minute OFF.
Indicates the presence of high voltage (1 kV or more). Take caution to guard against electric
shock.
Indicates that the surface becomes hot, and should not be touched with bare hands.
Indicates a non-specific general hazard. Follow the description given after this symbol or in
instruction manual.
Indicates that the main switch is ON.
Indicates that the main switch is OFF.
· Lamp housing for mercury burner
(U-LH100HG, U-LH100HGAPO)
· Illuminator
(BX3-URA, BX3-RFAS)
· ND filter high voltage]
(BX3-25ND6, BX3-25ND25)
· ND filter slider
(provided with BX3-RFAS)
· Power supply unit
(U-RFL-T)
Explanation
[Caution against
high temperature]
[Caution against
high voltage]