Olympus BX51 Instructions Manual - Page 27
Browse online or download pdf Instructions Manual for Microscope Olympus BX51. Olympus BX51 34 pages. System
Also for Olympus BX51: Operating Manual (17 pages), Instructions Manual (40 pages), Instructions Manual (37 pages), Operating Manual (11 pages)
5-2 Orthoscopic Observation
}In principle, polarized light enters the light path, parallel to the optical axis, to enable observation of the optical
characteristics of the specimen. Therefore, swing out the top lens of the condenser. Use 4X to 100X objectives.
1. When using the polarizing intermediate attachment U-CPA, pull out the Bertrand lens knob to remove the Bertrand
lens from the light path.
2. Insert the analyzer and perform observation.
# When the top lens is swung out, the aperture iris diaphragm and field iris diaphragm will not function
normally.
# If the aperture iris diaphragm is stopped down, the field of view may be restricted.
3. Insert a test plate (sensitive tint plate (U-TP530), quarter wave plate (U-TP137)) into the test plate slot. When fully
inserted, the plate is engaged. To remove the test plate from the light path, pull out to first click-stop position.
For details on other compensators, see the separate instruction manuals.
25