Olympus BX52-P Instructions Manual - Page 25
Browse online or download pdf Instructions Manual for Microscope Olympus BX52-P. Olympus BX52-P 34 pages. Polarizing microscope
Also for Olympus BX52-P: Instructions Manual (36 pages)
Fig. 31
23
3
Adjusting the Eyepiece Cross Lines
}To align the eyepiece cross lines and the vibration direction, an
orientation plate (B2-PJ) is required.
# Remove the test plate adapter and the compensator from the light
path.
1. Swing out the top lens of the condenser.
2. Place the orientation plate (B2-PJ) on the stage. Remove the analyzer
from the light path, and use the 4X objective to focus.
Orientation plate (B2-PJ)
(Fig. 31)