6 6
SPECIFICATIONS
Item
1. Polarizing
intermediate
attachment
(U-CPA and U-OPA)
2. Analyzer (U-AN360P) 360° dial rotation
3. Revolving nosepiece
(U-P4RE)
4. Stage (U-SRP)
27
Intermediate attachment conoscopic and
orthoscopic observation (U-CPA)
Field No.
Bertrand lens
Bertrand aperture stop
Bertrand lens knob position
for switch between orthoscopic
and conoscopic observation
Analyzer slot
Minimum scale reading: 0.1° (Vernier scale)
Type: Centerable quadruple
Attachable compensators: Quarter wave plate (U-TP137), sensitive tint plate (U-TP530) and all
kinds of compensators by means of test plate
Type: Polarizing rotatable stage with 3-point centering mechanism 360° horizontal rotation,
clamps at any desired position
360° scale (minimum division: 1°; minimum reading 6' by means of vernier scale)
45° click stop knob
· Stage clips (U-SCB2) attachable for holding specimen in place
· Attachable mechanical stage (U-FMP)
· Universal stage (ZEISS Co. manufacture) attachable
Specifications
orthoscopic observation (U-OPA)
Focusable
Fixed diaphragm
{
Pushed in position:
Pulled out position: ¦
Slot for rotatable analyzer (U-AN360P)
Intermediate attachment for
22