Olympus CK40 Instructions Manual - Page 34

Browse online or download pdf Instructions Manual for Microscope Olympus CK40. Olympus CK40 39 pages. Culture microscope

Olympus CK40 Instructions Manual
Trouble
5. Photomicrography
a) The picture is out of focus.
b) The edges are blurred.
c) The viewfinder image does
not look sharp.
d) A room window or fluorescent
lamp was also shot.
29
Cause
You are using the OM camera with the
shutter speed set faster than 1/2 sec.
You are using an achromatic objective.
This type of objective cannot bring edges
into sharp focus.
The correction collar of the LWDCDPlan
40XPL-6 is not adjusted.
Stray light entered through the eyepieces
or the camera's viewfinder.
Remedy
Use the ND filter and select a slower shut-
ter speed.
Use a Plan objective when necessary.
Adjust the correction collar according to
the thickness of the vessel bottom.
Cap both eyepieces and the camera's
viewfinder.
Page
24
18