Olympus CK40M FRAME Repair Manual - Page 13
Browse online or download pdf Repair Manual for Microscope Olympus CK40M FRAME. Olympus CK40M FRAME 38 pages. Compact inverted metallurgical microscope
CK40M
(4)
Mounting direction of voltage selector switch
Mounting direction
Make sure that the black cable is
connected to the 100 V (220 V)
side of the voltage selector switch.
B. DISASSEMBLY AND ASSEMBLY PROCEDURES
B-6
Voltage selector switch
Gray (100 V)
Brown (200 V)
Black (100 V/200 V)
Transformer ass'y
100/200
(AQ717800)
(AQ717900)