Olympus CK40M FRAME Repair Manual - Page 23
Browse online or download pdf Repair Manual for Microscope Olympus CK40M FRAME. Olympus CK40M FRAME 38 pages. Compact inverted metallurgical microscope
CK40M
2.
Preparation for adjustment
MS
Adjustment item
Absolute optical axis
Stage parallelism
CK4M-MA
illumination center
MA-AS center
Parfocality
Exit pupil center
Fine focus sensitivity
C. REPAIR PROCEDURE
Mirror
Stage insert plate
Eyepiece (jig)
Standard eyepiece
(KN0048, with Adapter-1)
Centering telescope
(KN0029)
Standard eyepiece
(KN0048, with Adapter-1)
Centering telescope
(KN0029)
Standard eyepiece
(KN0048, with Adapter-1)
FT-36 or focusing
telescope (-1) (KN0025)
Centering telescope
(KN0029)
Use WHK10X, WHK15X,
NCWHK10X or
GSWHK10X eyepiece.
Objective (jig)
Standard objective
(KN0041)
Gauge for checking stage
tilt alignment (SKN0003)
MDPlan80X objective
Standard objective
(KN0041)
MDPlan5X objective
Use MDPlan5X, 10X,
20X, 40X, 50X, 80X
objective (optional) or
MDAch10X, 20X, 50X
objective
C-2
Standard eyepiece
(KN0048, with
Adapter-1)
Centering telescope
(KN0029)
Standard tube
(BKN0003)
Standard objective
(KN0041)
Objective
(MDPlan5X)
(MDPlan80X)
Gauge for checking
stage tilt alignment
(SKN0003)
Jig and tool
Standard tube (BKN0003)
Thickness gauge 0.07 mm
(OT1949)
Standard tube (BKN0003)
Stage insert plate, mirror
Slide glass
Standard tube (BKN0003)
Stage insert plate, mirror
Standard tube (BKN0003)
Standard tube (BKN0003)
Use CH3-B145, CH3-TR45
or CH40-TBI observation
tube.