Olympus CK40M FRAME Repair Manual - Page 31

Browse online or download pdf Repair Manual for Microscope Olympus CK40M FRAME. Olympus CK40M FRAME 38 pages. Compact inverted metallurgical microscope

Olympus CK40M FRAME Repair Manual
CK40M
6.

MA-AS center adjustment

6-1
Set the following jigs as illustrated below to adjust the MA-AS center.
• Eyepiece ------KN0048: Standard eyepiece (with adapter-1)
KN0029: Centering telescope
• Objective------MDPlan80X
• Other ----------BKN0003: Standard tube, stage insert plate, mirror
The mirror side should
face down.
C. REPAIR PROCEDURE
Mirror
Stage insert plate
AS center adjustment position
ACU3X10SA (2 pcs.)
ANU4X6SA (1 pc.)
C-10
Standard eyepiece
(KN0048, with
Adapter-1)
Centering
telescope
(KN0029)
Standard tube
(BKN0003)
MDPlan80X