Olympus CX21 Instructions Manual - Page 18
Browse online or download pdf Instructions Manual for Microscope Olympus CX21. Olympus CX21 28 pages. Education microscope
Also for Olympus CX21: Maintenance Manual (25 pages)
SPECIFICATIONS
1. Optical system
2. Illumination
3. Focusing mechanism
4. Revolving nosepiece
5. Binocular observation tube
6. Stage
7. Condenser
8. Dimensions & weight
9. Operating environment
15
Item
UIS (Universal Infinity System) optical system
Built-in illumination system.
6 V, 20 W halogen bulb 6V20WHAL (PHILIPS 7388)
(Average life time: Approx. 100 hr. when used as directed)
100-120/220-240 V
Stage height adjustment mechanism.
Fine adjustment scale: 2.5 μm per graduation
Fine adjustment stroke: 0.3 mm per turn
Total stroke: 20 mm.
Pre-focusing knob provided, coarse adjustment knob tension adjustable.
Quadruple positions fixed (Front oriented)
Field number
Tube tilting angle
Interpupillary distance
adjustment range
Size
Movement range
Specimen holder
Type
N. A .
Aperture iris diaphragm
154(W) x 391(H) x 238(D) mm, approx. 5.9 kg
· Indoor use.
· Altitude: Max. 2000 meters
· Ambient temperature: 5° to 40°C (41° to 104° F)
· Maximum relative humidity: 80% for temperatures up to 31°C (88°F),
· Supply voltage fluctuations; Not to exceed ±10% of the normal voltage.
· Pollution degree: 2 (in accordance with IEC60664)
· Installation/Overvoltage category: II (in accordance with IEC60664)
Specifications
0.42/0.25A, 50/60 Hz
18
30°
48 to 75 mm
120 x 132 mm (with mechanical stage)
76 (X-axis) x 30 (Y-axis) mm
Holds a single specimen.
Abbe condenser (daylight filter detachable)
1.25 (when immersed in oil)
Built in
decreasing linearly through 70% at 34°C
(93°F), 60% at 37°C (99°F), to 50% relative
humidity at 40°C (104°F).