9-2 Assembly procedures
1
Removing the standard 10X eyepiece
a
2
Attaching the eyepiece micrometer
The size of the eyepiece micrometer which can be attached to eyepieces of this product is 24 mm in diameter and
TIP
1.5 mm in thickness.
4
28
TIP
1
Loosen the clamping screws a of the 10X eyepieces using the small
flathead screwdriver and remove the eyepieces.
1 1
To attach the eyepiece micrometer, loosen the clamping screw of the
right side eyepiece using a small flathead screwdriver and remove
b
the eyepiece.
Hold the eyepiece frame, loosen the reticle holder a and remove
2
the reticle holder a .
3
Insert the eyepiece micrometer b in the removed reticle holder a
3
with the display surface of the micrometer facing down.
a
NOTE
4
Screw in the reticle holder a which includes the eyepiece micrometer
b to the bottom of the eyepiece gently.
NOTE
4 5
Attach the eyepieces to the tube, and tighten the clamping screws.
The standard 10X eyepieces are clamped with screws.
Be careful not to leave dirt on the eyepiece micrometer b ,
since it will be noticeable during observation.
Be sure to screw in the reticle holder a completely until
it stops.