Olympus CX41 Instructions Manual - Page 14
Browse online or download pdf Instructions Manual for Microscope Olympus CX41. Olympus CX41 32 pages. System microscope
Also for Olympus CX41: Overview (7 pages), Instruction (32 pages)
Fig.11
Fig. 12
Fig. 13
11
3
Adjusting the Diopter
}When using the U-CTBI, align the white marking on the right eyepiece's
diopter adjustment ring scale with the index line.
1. Looking through the right eyepiece with your right eye, rotate the
coarse and fine adjustment knobs to bring the specimen into focus.
2. Looking through the left eyepiece with your left eye, turn the diopter
adjustment ring @ to focus on the specimen.
4
Using the Eye Shades
When Wearing Eyeglasses
Use with the eye shades in the normal, folded-down position. This will
prevent the eyeglasses from contacting and scratching the eyepieces..
When Not Wearing Eyeglasses
Extend the folded eye shades in the direction of the arrow for efficient
use of the eyeshades by preventing extraneous light from entering
between the eyepieces and eyes.
5
Light Intensity Ratio of Trinocular Tube U-CTR30-2
The U-CTR30 does not have the light path switching facility and the
light intensity ratio is fixed at binocular tubes (50%): TV/photography
(50%).
6
Photomicrography/TV Observation
1. Using the provided Allen wrench, fully loosen the clamping screw @
on the straight photo tube mount on the trinocular tube U-CTR30-2.
2. Attach the U-SPT straight photo tube ² or a TV adapter on the mount
and tighten the clamping screw @.
3. Attach the photomicrography system or TV camera.
(Fig. 11)
(Fig. 12)
(Fig. 13)