Olympus Fluoview FV1000 Overview - Page 18

Browse online or download pdf Overview for Microscope Olympus Fluoview FV1000. Olympus Fluoview FV1000 28 pages. Confocal microscope
Also for Olympus Fluoview FV1000: User Manual (42 pages), Short Instructions (7 pages), Short Instructions (7 pages), User Manual (9 pages), Manuallines (3 pages)

Olympus Fluoview FV1000 Overview
Laser Light Stimulation
The SIM scanner system combines the main scanner with a laser light stimulation scanner.
Control of the two independent beams enables simultaneous stimulation and imaging, to capture reactions
during stimulation.
Multi-stimulation software is used to continuously stimulate multiple points with laser light for simultaneous
imaging of the effects of stimulation on the cell.
FLIP—Fluorescence Loss in Photobleaching
Fluorescence loss in photobleaching (FLIP) combines imaging with continuous bleaching of a specific region to observe the diffusion of a
target protein within a cell. The changes in the image over time make it possible to observe the location of structural bodies that inhibit
the diffusion of the molecule.
Specimen: HeLa cell, GFP (free), 488 nm excitation (multi-argon laser)
Image acquisition time: 100 ms/ bleach time: 100 s continuously, 405 nm bleaching
FRAP—Fluorescence Recovery after Photobleaching
Exposure of fluorescent-labeled target proteins to strong laser light causes their fluorescence to fade locally. Fluorescence recovery after
photobleaching (FRAP) is used to observe the gradual recovery of fluorescence intensity caused by protein diffusion from the area
surrounding the bleached region. By examining the resulting images, it is possible to characterize the diffusion speed of the molecule,
and the speed of binding and release between the molecule and cell structures.
Example: Fluorescence recovery without interactions
If the protein can freely diffuse, the bleached region recovers
its fluorescence at a high speed due to Brownian motion.
Time
Specimen: Hippocampal neurons, Shank-GFP stain, 488 nm excitation (multi-argon laser)
Image acquisition time: 100 ms Bleach time: 80 ms, 488 nm excitation (Sapphire 488 laser)
Data courtesy of: Dr. Shigeo Okabe
Department of Anatomy and Cell Biology, Tokyo Medical and Dental University
Example: Fluorescence recovery with interactions
If the protein is strongly bound to a structure or forms part of a
large protein complex, the bleached region recovers its
fluorescence at a slower rate relative to the unbound state.
Time
17
3,000
2,800
2,600
2,400
2,200
2,000
1,800
1,600
1,400
1,200
1,000
800
600
400
200
0
0
10,000 20,000 30,000 40,000 50,000
60,000 70,000 80,000 90,000 100,000
Time (ms)
750
700
650
600
550
500
450
400
350
300
250
0
10,000
20,000
30,000
40,000
50,000
60,000
70,000
Time (ms)
80,000