Error Code List
The following table lists the error codes, their causes and remedies.
If the error recurs after applying the remedies described below, contact Olympus.
Error
Code
E01120
Command
E01130
Revolving nosepiece
Aperture iris diaphragm
E01150
Command
E01511
Command
E11214
Revolving nosepiece
E11218
Revolving nosepiece
E11271
Aperture iris diaphragm
E11272
Aperture iris diaphragm
E11278
Aperture iris diaphragm
37
Position
Phenomenon
Parameter error
Not connected error
Not connected error
Negative
acknowledgement
to a query
command
System Busy
Time-out error
Abnormality error
Origin error
Abnormality error
Abnormality error
Cause
The input string is errone-
ous.
Nosepiece connection fail-
ure.
The nosepiece is not con-
nected.
The connected nosepiece
is not the designated
model.
Cable wire disconnection.
A string was sent after a
query command.
The 1LOG IN command
was sent in the middle of
operation of a motorized
module.
The operation was not
completed in the specified
period of time.
Irrecoverable error.
Aperture iris diaphragm ini-
tialization after power ON
has not completed.
Irrecoverable error.
Irrecoverable error.
Remedy
Check the string used in the
transmitted command.
Check the nosepiece connec-
tion.
Connect the nosepiece.
Connect the designated nose-
piece.
Contact Olympus.
Re-check the command to be
sent.
The 1LOG IN command cannot
be issued while a motorized
module is being driven.
Issue the 1LOG IN command af-
ter the drive of the motorized
module has completed.
Revolving nosepiece: Check if
there is an obstacle for move-
ment.
Contact Olympus.
Switch on the power again.
Contact Olympus.
Contact Olympus.