Olympus MX61L Assembly Manual - Page 28

Browse online or download pdf Assembly Manual for Microscope Olympus MX61L. Olympus MX61L 49 pages. 200mm/300mm compatible semiconductor/fpd inspection microscopes

Olympus MX61L Assembly Manual

CENTRATION OF THE MERCURY/XENON BURNER

1

Turning the Burner On

Set the main switch of the power supply unit for mercury burner to " " (ON). The arc will stabilize in 3 to 5 minutes after
the burner is ignited.
# A discharge-type mercury or xenon burner may not turn on by the first try due to the characteristics of the burner.
If a burner does not turn on, set the main switch to "
to " " (ON) again.
# To avoid shortening the service life of the burner, do not turn a burner off in less than 15 minutes after turning it on.
# When turning on a mercury burner that has been turned off, wait for about 10 minutes after it is turned off,
because it cannot be turned on unless the mercury vapor in the burner has cooled down and liquefied.
# If the lamp housing is opened while the burner is on, the power supply stops to ensure safety. In this case, set the
main switch to "
lamp housing unless it has cooled down sufficiently.
# When resetting the burner hour counter, press and hold the reset button until the reading becomes "000.0".
2

Centering the Mercury Burner

}For centering of the xenon burner, refer to the instruction manual for the U-LH75XEAPO lamp housing.
@
A
B
C
D
25
" (OFF) and wait for more than 10 minutes before retrying to turn it on again. Do not open the
²
Fig. 35
" (OFF), wait for 5 to 10 seconds and set the main switch
}Set the main switch to " " (ON) and wait until the arc image stabilizes (for
5 to 10 minutes after ignition) before proceeding to the centering.
1. Remove an objective to make an idle position on the revolving nosepiece.
2. Select the BF light path using the light path selector knob.
3. Place a piece of white paper or similar object on the stage as the
specimen, and move the stage approximately to the focus plane of the
objective.
4. Open the aperture iris diaphragm.
5. Rotate the collector lens focus adjustment knob @ to project the arc
image on the white paper. (Fig. A)
If the arc image is not projected, rotate the burner centering knob ².
6. Rotate the lamp centering knob ² to move the arc image on the center
of the left (right) half of the field of view. (Fig. B)
7. Fit the tip of the Allen screwdriver into the mirror focus adjustment screw
³ (Fig. 36) on the rear of the lamp housing, and rotate the screw to focus
on the mirror arc image. (Fig. C)
8. Rotate the lamp centering knob ² so that the arc image and mirror arc
image are overlapped. (Fig. D)
}During observation, rotate the collector lens focus adjustment knob @ as
required to render the observed field of view uniform.
}It is not required to center the mercury burner further until the next time it
is replaced.
}If the collector lens focus adjustment knob @ is located to far to be
manipulated, fit the U-CLA extension handle (optional) shown below into
the knob @ and manipulate them together.
(Figs. 35 & 36)