Olympus SZX10 Instructions Manual - Page 30

Browse online or download pdf Instructions Manual for Microscope Olympus SZX10. Olympus SZX10 36 pages. Research high-class stereo microscope
Also for Olympus SZX10: Overview (24 pages), Instructions Manual (40 pages)

Olympus SZX10 Instructions Manual

9-4 BX Stage Adapter Type 1 SZX-STAD1

}This adapter is for installation of a U-SRG or U-SRP rotary stage on the SZX2 standard base or a SZX2/SZX series
illuminator stand. When the U-SRP rotary stage is used together with the U-FMP mechanical stage, X-Y directional
movement becomes possible, which is convenient for framing during photomicrography.
To cover the height of the stage adapter (about 44 mm), we recommend combination with the SZH-P400 auxiliary pillar
(together with the SZX-R drop prevention collar).
1
Mountable Bases, Restrictions
Standard base
SZX2-ST/SZX-ST
· Transmitted illuminator stand
SZX2-ILLK/SZX-ILLK
· High-class illuminator stand
SZX2-ILLB/SZX-ILLB2
· Brightfield/darkfield transmitted
illuminator stand
SZX2-ILLD/SZX-ILLD2
26
Base
The same restrictions apply regardless of whether the stage adapter is used
or not. (Refer to the instruction manual for the SZX2/SZX illuminator stand.)
Note that the field illuminated by the transmitted light is limited depending
on the diameter of the opening of the stage plate in use.
# Darkfield observation is not possible using the SZX2-ILLD or SZX-ILLD2.
# The illumination intensity may be reduced when a frosted filter is used.
Applicable Objectives
0.5X to 2X
Restrictions
None