Olympus SZX16 Instructions Manual - Page 17
Browse online or download pdf Instructions Manual for Microscope Olympus SZX16. Olympus SZX16 40 pages. Research high-class stereo microscope
Also for Olympus SZX16: Overview (24 pages), Instructions Manual (36 pages), Instructions Manual (28 pages)
4-4 Camera Observation and Photomicrography
A digital camera unit can be mounted on the straight tube mount of the trinocular tube by means of the camera
adapter and/or camera mount adapter*.
* The camera mount adapter is not necessary if a camera adapter equipped with a camera mount is used.
For details, please also read the instruction manuals for the camera adapter and digital camera.
1
Selecting the Camera Adapter Magnification
Set the magnification of the camera adapter according to the size of the CCD in the digital camera.
(Example) The following figures show the camera observation/photomicrography areas when the WHN10X
eyepieces with FN 22 are used.
When the 0.5X camera adapter is used
b
Downloaded from
ManualsNet.com
a
c
search engine
1 in. CCD
2/3 in. CCD
1/2 in. CCD
WHN10X
FN 22
When the 1X camera adapter is used
2
Mounting the Camera Adapter
1
Using the Allen screwdriver, fully loosen the straight tube clamping
screw a of the straight tube mount on the top of the trinocular tube.
2
Fit the round dovetail b of the camera adapter into the straight tube
mount of the trinocular tube, and tighten the clamping screw a .
3
Mount the camera on the camera adapter. A camera mount adapter
may be required with certain camera adapters.
3
Selecting the Camera Light Path
Pull out the light path selector knob c to select the Binocular 50%/
Camera 50% light path setting for the SZX2-TR30/TTR or the Camera
100% light path setting for the SZX2-TR30PT/TTRPT.
SZX16
13