Olympus U-EXBABG Instructions Manual - Page 38
Browse online or download pdf Instructions Manual for Microscope Olympus U-EXBABG. Olympus U-EXBABG 42 pages.
³
Fig. 17
ƒ
|
Fig. 18
8
Setting Up the Power Supply Unit
Cables and cords are vulnerable when bent or twisted. Never sub-
@
ject them to excessive force.
Make sure that the main switch is set to " " (OFF) before connecting
the power cord.
²
Always use the power cord provided by Olympus. If no power cord
is provided with the microscope, please select the proper power
cord by referring to chapter "PROPER SELECTION OF THE POWER
SUPPLY CORD" at the end of this instruction manual.
1. Verify that the voltage and frequency of the input power supply match the
requirements inscribed on the rating plate @.
(The 100 V system can be used with voltages in the 100 to 120 V range
and 200 V system can be used with voltages in the 220 to 240 V range,
both with frequencies of 50 or 60 Hz.
2. Securely plug the burner socket connection cord into the power supply
unit's connector ².
3. Plug the power supply unit's power cord into its power input connector
³, then plug the power plug | into the wall power outlet ƒ.
Be sure to supply power from a grounded, 3-conductor power outlet
using the proper power cord. If the power outlet is not grounded
properly, Olympus can no longer warrant the electrical safety perfor-
mance of the equipment.
(Figs. 17 & 18)
35