Olympus CelonLab ENT Instruction - Page 6
Browse online or download pdf Instruction for Medical Equipment Olympus CelonLab ENT. Olympus CelonLab ENT 10 pages. Safety check
Fig. 3:
Sample of the front panel of the power control unit showing the HF-outputs.
4. Valid resistance range
No.
Test equipment:
Test accessories: Power cord
200
Connect the unit via the power cord with a rated supply voltage. Connect the footswitch with the unit.
210
Connect the unit via the connection cables at the HF-output 1 and 2 (see fig. 3) with the electrosurgical
analyzer according to the analyzer's instructions for use. Set the power level of the unit to "12" and
measure the following parameter:
Attention: Start check procedure, if not already done, after setting the power level to "12"and
make sure to release the footswitch before starting with a new check procedure.
Short circuit detection
The following signals and warnings have to occur at an output load of 10 Ω (unit shut off):
220
The blue LED (signal lamp "power output") is dark.
-
The red LED (signal lamp "resistance too low") is blinking.
-
An alarm signal (pulsed tone) is audible.
-
The following signals have to occur with a load resistor of 25 Ω (normal function of the unit):
230
The blue LED (signal lamp "power output") glows.
-
The red LED (signal lamp "resistance too low") is dark.
-
A continuous signal is audible.
-
© Celon AG
,
Issue date: 01.09.2008
SCI.991007-1.4
Safety Check
Instruction.doc
Safety Check Instruction
Electrosurgical analyzer (Example: QA-ES, Metron)
Connection cables, HF-output (banana) to the electrosurgical analyzer
Footswitch
Procedure
HF-output 2
HF-output 1
Page 6 / 9