Cautions and recommended electromagnetic environment regarding portable and
mobile RF communications equipment such as a cellular phones
Immunity test
IEC 60601-1-2
test level
Conducted RF
3 Vrms
IEC 61000-4-6
(150 kHz to 80 MHz)
Radiated RF
3 V/m
IEC 61000-4-3
(80 MHz to 2.5 GHz)
NOTES
• Where "P" is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and "d" is the recommended separation
distance in meters (m).
• This instrument complies with the requirements of
IEC 60601-1-2: 2001 and IEC60601-1-2: 2007.
However, under electromagnetic environment that
exceeds its noise level, electromagnetic interference
may occur to this instrument.
Recommended separation distance between portable and mobile RF communications
equipment and this instrument
Rated maximum output
power of transmitter
P (W)
0.01
0.1
1
10
100
NOTE
The guidance may not apply in some situations.
Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Portable and mobile RF communications equipment
such as cellular phones should be used no closer to any
part of this instrument, including cables than the
recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Compliance level
3 V (V
)
1
3 V/m (E
)
1
Separation distance according to frequency of transmitter (m)
(calculated as V
=3 and E
1
1
150 kHz to 80 MHz
d=1.2
0.12
0.38
1.2
3.8
12
Guidance
Formula for recommended separation distance
(V
=E
=3 according to the compliance level)
1
1
3.5
d=[
]
V
1
3.5
d=[
]
E
1
7
d=[
]
E
1
• Electromagnetic interference may occur to this
instrument near a high-frequency electrosurgical
equipment and/or other equipment marked with the
following symbol:
=3)
80 MHz to 800 MHz
d=1.2
0.12
0.38
1.2
3.8
12
80 MHz – 800 MHz
800 MHz – 2.5 GHz
800 MHz to 2.5 GHz
d=2.3
0.23
0.73
2.3
7.3
23
EMC Information