Olympus CIX100 Systems Setup Manual - Page 26
Browse online or download pdf Systems Setup Manual for Laboratory Equipment Olympus CIX100. Olympus CIX100 38 pages. Cleanliness inspector system
Also for Olympus CIX100: Installation Manual (14 pages), Instructions Manual (46 pages)
OLYMPUS Cleanliness Inspector CIX100
10 Troubleshooting
Solution
System Setup Guide
1. Perform the calibration processes
and
[Camera-to-Stage
home page of the software opens the dialog box that contains the
calibration processes.
2.
[Camera-To-Stage
Rotation] calibration process: If the rotation angle is
more than 1°, contact Customer Service:
Phone: (+ 49) 251-79800-6444, fax: (+ 49) 251-79800-6445
3. When both calibration processes are completed successfully, perform
the
[Check
System] workflow because the calibration processes
invalidate the existing system check.
Then perform the
[Inspect
images are assembled correctly.
4. If there are still problems with the assembly of the images, please
contact Customer Service:
Phone: (+ 49) 251-79800-6444, fax: (+ 49) 251-79800-6445
26
When the images have been assembled this
way, it indicates that the
Rotation] calibration is no longer valid.
[Manual Magnification
Rotation]. Clicking the [Calibration] button on the
Sample] workflow and check if the recorded
[Camera-to-Stage
Calibration]