Olympus CIX100 Instructions Manual - Page 15
Browse online or download pdf Instructions Manual for Laboratory Equipment Olympus CIX100. Olympus CIX100 46 pages. Cleanliness inspector system
Also for Olympus CIX100: Systems Setup Manual (38 pages), Installation Manual (14 pages)
5
Calibrating Shading Correction for a new sample type
Please make sure to log into the CIX ASW software as administrator. Then you have
full access to the system configuration on the start page of your CIX ASW software.
Insert a clean cellulose membrane to your system with one of the sample holders.
Calibrate the stage limits with this sample as described in the CIX ASW software's
help.
1. On the start page, click the [Calibration] button.
2. In the [Calibrations] dialog box, select [Shading Correction] and click
[Calibrate].
In the following dialog box, you find a list of observation methods. In
l
your CIX100 system, each observation method represents a list of
software settings optimized for the analysis under specific optical
conditions. Each sample type requires its own observation method. If
your system is equipped with a real color slider, two observation
methods will have to be calibrated for each sample type for optimal
results. The observation method with real color slider is named after
the particle type appended with "in Real Color Mode" and is always
listed below the observation method for the respective particle type
without real color slider.
Sample Holders
15