ASSEMBLY
3-1 Assembly Diagram
}For the assembly of the SZX microscope, refer to the separate instruction manual for the SZX microscope.
# Remove dirt and dust from each part before assembly, and assemble the parts carefully by following the numbers
shown in the figure.
# When the SZX-RFA is combined with another intermediate attachment, some part of image may be cut off and the
performance may be degraded. Also note that the SZX-RFA cannot be combined with the SZX-ILLC.
In case the SZX-RFA is combined with another intermediate attachment, always position the intermediate attach-
ment above the SZX-RFA. If the intermediate attachment is located below the SZX-RFA, it cannot exhibit full illumina-
tion performance.
# When the SZX-BS beam splitter is used together with this attachment, it is not possible to attach the SZX-SLR single
lens reflex adapter onto the left side.
The SZX-BS manufactured prior to February 1998 has a light path switch knob that is too long and interferes with the
SZX-RFA. Contact your local Olympus representative for assistance.
Coaxial fluorescence
illuminator
SZX-RFA
Microscope
SZX12
3
Collector lens
IX-UCLHG/XE
SZX observation tube
SZX-BI30
SZX-TBI
SZX-TR30
Power supply unit
(for 100W mercury burner)
Mirror unit
SZX-MGFP
SZX-MGFPA
SZX-MG
SZX-MIY
Mercury burner
USH102D
Lamp housing
U-ULH
Power cord
Socket
U-ULS100HG