Olympus U-AFA2M-VIS Instructions Manual - Page 12
Browse online or download pdf Instructions Manual for Laboratory Equipment Olympus U-AFA2M-VIS. Olympus U-AFA2M-VIS 24 pages. Active autofocusing unit
3-2 Detailed Assembly Procedures
1
Installing the U-AFA2M Active AF Unit
²
@
Fig. 2
³
Fig. 3
8
CAUTION
1. Fully loosen the observation tube clamping screw @ using an Allen
# The centering of the AF unit should be adjusted before mounting it
2. Attach the standard straight tube ³ on the observation tube mount on
3. Engage the objective with the highest magnification of the ones being
4. Set the observation path of the vertical illuminator to reflected light
5. Slide the top of the CT to focus on the objective pupil.
6. Loosen the clamping screws in the four holes 5 on the top of the
|
5
7. Remove the standard straight tube and CT, and block the 4 holes
Be sure to adjust centering of the AF unit.
To ensure safety, be sure to unplug the AF cable during installation
and adjustment of the AF unit.
screwdriver, attach the circular dovetail on the bottom of the active AF
unit ², and then tighten the clamping screw @ again.
on the observation tube.
the AF unit, then insert the scaled CT | into the standard straight tube.
Rotate the helicoid on the top of the CT to focus on the cross scale.
used in the light path.
brightfield (BF).
sensor unit, align the center of the objective pupil with the cross of the
CT (within one scale in both up/down and left/right directions), then
tighten the clamping screws firmly.
with the provided contamination stickers.
(Figs. 2 & 3)