Olympus U-LLGAD Instruction Manual - Page 6

Browse online or download pdf Instruction Manual for Laboratory Equipment Olympus U-LLGAD. Olympus U-LLGAD 20 pages. Light guide illumination system

Olympus U-LLGAD Instruction Manual
1

Getting Ready

1)
This Instruction Manual pertains only to the light source unit. Read also the Instruction Manual of the microscope to
be used together with this equipment so that you can understand the comprehensive use procedures.
2)
This equipment is a precision unit. Handle it carefully to avoid sudden and severe impact. This equipment is not
waterproof.
3)
Do not use this equipment where it is subjected to direct sunlight, dust, steam or vibration. For operating
environment, refer to [5 Specifications] on page 11.
4)
Highly flammable organic solvent, e.g. dehydrated alcohol, etc. may cause fire. Keep it away from the light source
unit while the light source is turned on or immediately after the light source is turned off.
5)
When the liquid light guide is broken, solvent leaks out. Though the solvent is not severely toxic, wash your hands
and/or eyes thoroughly with water if they got in touch with the solvent.
6)
Do not twist the liquid light guide excessively to prevent it from being damaged. Also do not pull it with more than
3kg tension.
7)
Be sure to set the power switch to \ (OFF) and unplug the power cord from the outlet and remove the liquid light
guide before carrying the equipment.
8)
Be careful not to touch around the ventilation opening area while this equipment is turned on or immediately after it
has been turned off. As this ventilation opening area is heated, this causes burns.
9)
Do not repeat turning ON/OFF the lamp more than necessary. The life of the lamp may be shortened.
After the lamp is turned ON, wait for 2 minutes before turning OFF the lamp.
After the lamp is turned OFF, wait for 10 minutes before turning ON the lamp.
2
Maintenance and Storage
1)
Clean periodically (every 100 hours as a guide). Before cleaning, unplug the power plug from the outlet and wipe it
with a fluff-free soft dry cloth.
2)
If the light source unit and the liquid light guide are very dirty, wipe them with the soft cloth slightly containing the
diluted detergent. If the edge surface of the liquid light guide is dusty, blow the dust off with the clean air. If the dust
cannot be removed, clean it off with the cotton swab or the lens paper containing the alcohol.
3)
Vacuum the dust around the ventilation openings on the bottom and side of the light source with a vacuum-cleaner
periodically (every 6 months as a guide). Using the dusty equipment continuously may result in a fire or failure.
4)
Never disassemble/remodel anything other than those described in this Instruction Manual. This may cause the
performance deterioration or incidents.
5)
If you don't use the equipment, be sure to turn \ (OFF) the power switch and unplug the power cord from the outlet.
Then, after confirming that the equipment is cooled down completely, store it and put the dust prevention cover over
it.
6)
When the lamp replacement instruction of the equipment appears (when the lamp "lighting time" blinks), install a
new lamp. (See page 9.) The mercury lamp, different from the fluorescence, contains a high pressure gas within the
tube. If you keep using it beyond the lifetime extremely, the glass tube may burst due to the accumulated strain. The
used mercury lamp must be disposed of as industrial waste. If you cannot dispose of it by yourself, bring it to the
Olympus sales office.
7)
Before disposing of this product, be sure to follow the regulations and rules of your local government.
In case that the lamp bursts
Should the lamp burst, follow the procedures below.
• Should the lamp burst, unplug the power cord from the outlet. Leave that place and ventilate for at least 30 minutes.
• After the lamp and the lamp house have cooled down, collect the remaining mercury with packing tape, paper or
dropper, etc.
• The collected mercury and tools used must be sealed completely in a nonmetallic container, and ask the waste
disposer to dispose them.
• Should you have inhaled mercury steam, consult the doctor immediately and follow his/her instructions.
2