Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level
Guidance and manufacturer's declaration – electromagnetic immunity
The MODEL ABP801 is intended for use in the electromagnetic environment specified below. The customer or the
user of the MODEL ABP801 should assure that it is used in such an environment.
Immunity
IEC
test
level
Conducted RF
3 Vrms
IEC
150 kHz to
61000-4-6
80 MHz
Radiated RF
3 V/m
IEC
80 MHz to
61000-4-3
2.5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically
with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should
be considered. If the measured field strength in the location in which the MODEL ABP801 is used exceeds the
<5 % UT
(>95 % dip in U ) T
for 5s
3 A/m
60601
test
Compliance
level
3 Vrms
3 V/m
<5 % UT
(>95 % dip in U ) T
for 5s
3 A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.
Electromagnetic environment-guidance
Portable and mobile RF communications equipment
should not be used close to any part of the MODEL
ABP801, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter
recommended separation distance
80 MHz to 800 MHz
80 MHz to 2,5 MHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey, should
be less than the compliance level in each frequency
range b.
Interference may occur in the vicinity of equipment
marked with the following symbol:
12