ABB AnalyzeIT ZMT Series Supplementary Operating Instructions - Page 6
Browse online or download pdf Supplementary Operating Instructions for Measuring Instruments ABB AnalyzeIT ZMT Series. ABB AnalyzeIT ZMT Series 20 pages. Zirconia oxygen analyzer systems
5 SETTING UP
For all aspects other than serial data transmission the
transmitter is set up as shown in the Operating Instructions
(IM/ZMT). Unless otherwise requested, the instrument is
despatched
with a transmission rate of 9600 baud and
transmission line termination resistors linked-out. If these
settings are unsuitable, refer to Section 5.1.
5.1 Termination Resistors
and Baud Rate Settings – Fig. 5.1
5.1.1 Termination Resistors
For long transmission lines, termination resistors are
required on the last ZMT Analyzer in the chain and at the host
computer/computer terminal. Under normal operating
conditions the resistors are required at the last ZMT receive
inputs only – see Fig. 4.1. The analyzer's resistors are
selected using plug-in links – see Fig. 5.1.
Switch off the supply and gain access to the Serial Output
board in Output Module 3 – see IM/ZMT, Section 6.2 . Set the
termination resistor links as shown in Fig. 5.1.
5.1.2 Hardware Baud Rate Setting
To set the Baud rate to the required speed, set the plug-in link
as shown in Fig. 5.1
6
1
IN
7
12
6
1
OUT
7
12
Termination Resistors
Fig. 5.1 Termination Resistor and Baud Rate Links
4
1
16
1200
2400
4800
9600
8
9
Baud Rate Links
6 PROTOCOL
The protocol used is based on ANSI-X3.28-1976-2.5-A4 and
is used for master (host computer) to slave (ZMT Analyzer)
systems. This is the
recommended protocol for use with
such as ABB Kent-Taylor PC30. The
supervisory systems
Protocol is:
Start transmission (STX) – Command – Identification...
...End transmission (ETX) – see Figs. 8.1 to 8.6.