Digiray Fire CR Maintenance Manual - Page 7
Browse online or download pdf Maintenance Manual for Dental equipment Digiray Fire CR. Digiray Fire CR 11 pages. Veterinary dental cassette kit
Maintenance Guide
Veterinary Dental Cassette Kit
2.1. Service Life
Imaging plates in the following condition cannot be used.
Scratches on the surface of the protective layer and they appear on the image.
The phosphor layer has peeled away from the support layer.
2.2. Imaging Plate Cleaning Tools
Moving plate extracting tool
Imaging plate cleaner
Use ethanol based cleaner : ethanol > 99.5%
Soft lint free dry cotton (nonabrasive cloth)
2.3. Imaging Plate Cleaning Process
2.3.1. Moving Plate Pull Out
7
TM-723-EN