Siemens SIREMOBIL Iso-C Service Instructions Manual - Page 16

Browse online or download pdf Service Instructions Manual for Industrial Equipment Siemens SIREMOBIL Iso-C. Siemens SIREMOBIL Iso-C 16 pages.
Also for Siemens SIREMOBIL Iso-C: Installation And Setting Instructions (4 pages), Installation And Setting Instructions Manual (6 pages), Maintenance Instruction (46 pages), Installation Instructions Manual (6 pages)

Siemens SIREMOBIL Iso-C Service Instructions Manual
16
2
Fig. 32
Reattach the cover plate (1/Fig. 32).
Secure the cable to the switch board with cable clamps.
Reattach the lower telescope cover (1/Fig. 33).
Reattach the cover (2/Fig. 33) at the Velcro closures.
Reconnect plug X1 to the cover with the brake switches.
Reconnect the protective conductor to the cover with the brake switches.
Remount the cover with the brake switches to the horizontal carriage.

Concluding work

Check the horizontal lift for ease of movement.
Check the angulation for ease of movement.
After completing all service work and reattaching the covers, perform the protective
conductor test in accordance with ARTD-002.731.17.
Perform a function test on the SIREMOBIL.
TD SD 24 / Arnold
TD SD 24 / Groß
SMS Iselin / O'Donnell
SIREMOBIL Iso-C
Replacing the horizontal lift brake
1
S3
S4
SPR2-230.076.01
Rev. 01
2
Fig. 33
Page 16 of 16
02/99
TD SD 24
1
Siemens AG
Medical Engineering
1