38 Maintenance, Operating Value / Functional Inspection
4-
4
Maintenance, Operating Value / Functional Inspec-
tion
4.1
Maintenance
PMP
Cleaning the system
•
Clean the entire system:
- Visible cables
- Exterior surfaces
- Contact surface of wheels
- Interior space
PMP
System ventilation
•
Clean the ventilation slots of the SIREMOBIL stand and monitor cart.
PMP
Memoskop ventilation
•
Clean the ventilation slots on the Memoskop.
•
Check that the fan is functioning properly.
4.1.1
DICOM bridge
PMP
Ventilation
•
Clean the ventilation slots of the DICOM bridge.
•
Check that the fan is functioning properly.
4.1.2
Codonics printer
PMP
Ventilation
•
Clean the ventilation slots of the printer and Framegrabber.
•
Check that the fan is functioning properly.
PMP
Cleaning
NOTE
SIREMOBIL Iso-C
According to the EP1660 user manual, it is the responsibility of the
customer to clean the thermal print head and the mechanics for
film transport. The interval between cleanings is also specified in
the user manual. Refer to the maintenance chapter for the Codon-
ics printer. The annual cleaning of the printer described here is not
a substitute for regular maintenance by the customer. In the event
that malfunctions occur or image quality worsens due to insuffi-
cient cleaning, please refer the customer to the maintenance activ-
ities/interval described in the Codonics user manual.
SPR2-230.101.02.06.02
07.05
Page 38 of 46
CS PS 24
Siemens AG
Medical Solutions