Concluding steps

Reinstall all the covers.
Perform the protective conductor measurement as specified in ARTD-002.731.17.
Observe the information on protective conductor testing in these instructions.
Perform the protective conductor test according to ARTD-002.731.17.
Perform partial acceptance tests in accordance with country-specific regulations.
Changes to previous version
Entire instructions extended to SIREMOBIL Compact L as from serial number 3000.
Safety information newly inserted.
Required measuring equipment and auxillary devices revised.
Check of the programmed values revised.
Check of the accuracy of the area dose product measuring instrument revised.
SIREMOBIL Iso-C
Area dose product measurement device
The protective conductor resistance may not exceed 0.2 Ohms.
SPR2-230.033.04
Rev. 03
06.05