Siemens SITRANS F M Verificator Operating Instructions Manual - Page 13
Browse online or download pdf Operating Instructions Manual for Industrial Equipment Siemens SITRANS F M Verificator. Siemens SITRANS F M Verificator 38 pages.
3.4
Verificator software - sensor - magnetism
Test of magnetic properties in the sensor
The bar indicates how the test is progressing. Press ESC to cancel the test.
T3:3 Magnetism test
0% -##
Press ESC to cancel
Test of the magnetic properties finished
Press ESC to cancel the verification.
T3:3 Magnetism test
0% -################
Press ESC to cancel
Is only displayed if the SENSORPROM
Press GO to store magnetic properties data in the SENSORPROM
reference data for future verifications.
Always make sure that the sensor is OK before pressing GO.
Press ESC to cancel the verification.
Update SENSORPROM®
(see manual)
Press GO to update
Press ESC to cancel
Is only displayed if the previous display was accepted by pressing GO.
SENSORPROM
Restarting transmitter
Update SENSORPROM®
Please wait...
Is only displayed if the measured „fingerprint" is out of range.
Could be caused by a defective Verificator. See "On site trouble shooting".
E3:
Sensor:
Press ESC to cancel
SITRANS F M Verificator
Operating Instructions, 07/2019, A5E45134218-AB
-100%
-100%
®
unit is updated with the data about the magnetic properties.
®
Sensor Failed
Selftest Fail
3.4 Verificator software - sensor - magnetism
unit installed does not contain fingerprint data.
Operating (hardware)
unit. This data is used as
®
13