Siemens SITRANS L Series Compact Operating Instructions - Page 28
Browse online or download pdf Compact Operating Instructions for Switch Siemens SITRANS L Series. Siemens SITRANS L Series 45 pages. Capacitance transmitters
Also for Siemens SITRANS L Series: Operating Instructions Manual (33 pages), Operating Instructions Manual (50 pages), Operating Instructions Manual (44 pages)
Commissioning
6.2 Local commissioning
6.2.6.3
Handheld programmer: function keys in RUN mode
Key
6.2.6.4
Accessing a parameter
Note
• The following instructions apply when using the Handheld Programmer.
• Do not use the Handheld Programmer at the same time as SIMATIC PDM, or erratic
operation may result.
• You do not need to key in initial zeros when entering a parameter number: for example,
for P005, key in 5.
28
RUN mode
mA output value displayed in auxiliary reading field.
Internal enclosure temperature displayed in auxiliary reading field.
Parameter for auxiliary readings
Press
to display the auxiliary reading field when in RUN mode.
Displays the value representing Echo Confidence (P805).
Toggle between Units and % on reading display.
Initiate and complete PROGRAM mode access.
Distance displayed in auxiliary reading field.
SITRANS Probe LR with mA/HART
Compact Operating Instructions, 05/2019, A5E46248119-AC