III.
Menu Options
Foreign Language Option
The menu options can be adjusted to read in German,
French, or Spanish. To change the language, start the unit as
you normally would. When the ACD screen appears,
A. C. D. I
Copyright 2003
press the SELECT button. This will bring up the following
screen:
SELECT to change
Press the SELECT button to choose the preferred language.
When the language is displayed, press the POWER button to
continue with the start up sequence.
Flow-Rate Select Option
The CAL 2000 LT is shipped from the factory set at either 0.5
LPM or 1.0 LPM depending on what was specified when the
unit was ordered. To change this in the field the following
screen can be used.
Select Flow Rate
0.5 LPM 1.0 LPM
Note that the CAL 2000 LT cells read this setting on the in-
strument and will work correctly at either flow-rate.
Page 11
English
The Calibration Date refers to when the instrument itself
was last calibrated at the Factory. It does not refer to the
calibration date of the generating cells, which are field
replaceable.
The following status screens indicate the remaining cell
life, battery status and flow-rate.
Cell Life
009 Hrs 59 Mns
Battery Status
100% remaining
Flow: 0.5 lpm
Immediately after the flow rate screen appears, the ppm
screen appears.
Output: 5 ppm