DigiVac SNAP Operational Manual - Page 15
Browse online or download pdf Operational Manual for Controller DigiVac SNAP. DigiVac SNAP 17 pages. Vacuum control and pressure regulation in one
Section 7. Specifications
Valves
Range of control
Accuracy, control
Time to converge within 5%
after disturbance
Integral sensor
Integral Sensor Accuracy
Integral Sensor Range
Vacuum Path Orifice
Minimum Bleed Valve
Orifice
Power
Dual Vacuum Control
External Sensor
Output
Recipes
Enclosure
Certifications
©
2021 DigiVac Company®Allrights reserved. | 1020 Campus Drive, Morganville,NJ 07751 | (732) 765-0900
Wetted materials: 304 Stainless Steel, Viton, and PTFE (teflon)
2 Torr to 775 Torr
+/- 5% of reading
<30 seconds
SEN-775i
+/- 2 Torr
0.5-775 Torr
Minimum orifice of 20mm
¼" or 6.35mm hose barb I.D.
100-240VAC 50/60Hz
Proportional throttle and proportional bleed control (vent to
atmosphere) delivered from an integral dual valve module per channel
Upgrade to a 10 torr capacitance manometer for improved accuracy
Ability to be controlled remotely via USB or analog in/out for integration
into larger systems
Program ramps and holds vacuum at different duration and vacuum
levels
9 ¼" width, 4 ¼" depth, 4 ¾" height
CE, UL, CSA, (planned for 1H 2022) RoHS
15