4. Event Log
By default the personal monitor stores the last twenty five (25) alarm events. The system stores events
by first in first out, i.e. the 26
downloaded using the Crowcon IR Link. For each alarm event the personal monitor records the
following:
The personal monitor serial number
Bump Test (Yes or No)
Life remaining on the personal monitor
Number of self tests performed
Number of events
Alarm Condition (Low, High, or OL)
Specific event date and time
Peak gas concentration in ppm or %.
5. Bump Test Interval
Using the IR Link, personal monitors can be programmed to alert the user if a bump test is due. This
interval can be set anywhere from 1 to 365 days. *Note the default is to have no bump interval
programmed.
If a personal monitor is due for a bump test, the personal monitor will flash 'buP' on the screen and if
configured to do so will emit alternating LED flashes (left and right), and the 'TEST' icon will remain
even after a button press. This alert can be cleared by either placing the personal monitor in a C-Test
Station or manually performing a Bump Test. To manually perform a Bump Test, affix the flow plate to
the sensor aperture and ensure the correct level of gas is connected via a fixed flow regulator (0.5 l/m).
An exhaust pipe should be used into a safe area where no personal are present. Press the button down
once, the word "GAS" will show on the display while the TEST icon flashes. The personal monitor will
wait for 45 seconds for the target gas to be applied, or a button press to skip the bump test. If the
Crowcon Clip or Crowcon Clip+ is bumped while showing "GAS", then it will record as a bump test in
the event log instead of as an exposure. If no gas is applied, it will return to the normal screen and will
not record anything in the event log.
Oxygen Calibration
Crowcon recommends users of the O
before using each day.
The Crowcon Clip Oxygen (O
personal monitor every 30 days. The user will be prompted by the screen flashing CAL, please see
calibration instruction below:
Only perform O2 calibration in clean air (20.9% Oxygen) that is free of hazardous gases.
Calibration Instructions
Issue 4 – December 2016
th
event will replace the first event and so on. This information can be
Crowcon Clip to bump test the personal monitor
2
) personal monitor factory default will prompt the user to calibrate the
2