aeolus UC-1901M User Manual - Page 10

Browse online or download pdf User Manual for Pet Care Product aeolus UC-1901M. aeolus UC-1901M 10 pages. Professional multifunctional veterinary intensive care unit

aeolus UC-1901M User Manual
4. CO
Sensor
2
If CO
sensor alarm sounds,
2
please check the emergen-
cy vent first.
If emergency vent isn't open,
please open it manually by
se�ng Emergency Vent to
"ON" and observe if carbon
dioxide concentra�on drops.

Non-alarming components

1. Nebulizer
a) Please check if foreign objects or drug residue have blocked the medicine cup or nebulizer pipeline.
b) Open and remove the top cover on the ICU compartment to expose internal components.
c) Please find the nebulizer pump and check if it's damaged; replace it when needed.
Note: There is a backup nebulizer pump inside, please double check and diagnose which one is
broken before replacing.
If emergency vent is open,
please
check
if
the
ven�la�on fans on top of the
compartment are working
properly.
If the ven�la�on fans are working properly
please open the grid at the innermost top
corner of the compartment to check and
replace the carbon dioxide sensor if
needed.
16
2. Lighting Systems
Open the grid at the top corner nearest to the
door hinge.
Locate faulty light component and replace it
when needed.
4. Fans
Open the top cover of the ICU unit; locate the
fan assembly as indicated by the red arrow in
Figure below.
Replace the fans when necessary.
Note: The PTC heating assemblies will need to
be removed before replacing the fans.

Maintenance And Cleaning

Both inside and outside of the machine can be wiped clean regularly with a mild cleaning agent that
is non-abrasive, such as "Aseptic Plus" clinical grade disinfectant.
Clean the control panel with a mild household cleaner and non-abrasive cloth or sponge at least one
time each week.
All inlet vents should be vacuumed regularly to remove any dust; they can be wiped clean with a soft
cloth using a mild cleaning agent. We suggest the filters are removed and washed clean regularly,
then allowed to dry before replacing.
In between usages, the compartment should be cleaned with a soft cloth or sponge using a mild
cleaning agent; then turn the ultra-violet sterilization on for at least 3 minutes to disinfect the whole
unit.
Regular testing of the emergency vent should be done between usages; the unit can simply be
powered off to check if the vent opens.
3. Negative Ion Generator
Open and remove the top cover on ICU
compartment to expose internal components.
Please find the negative ion generator, check if
it's broken.
5. Emergency Vent
Open and remove the protective cover on the
emergency ventilation fan on the back of the
unit, as indicated in Figure below.
Replace the entire emergency vent assembly.
17