aeolus UC-1801 Quick Start Manual - Page 4

Browse online or download pdf Quick Start Manual for Accessories aeolus UC-1801. aeolus UC-1801 12 pages. Professional vet equipments supplier

aeolus UC-1801 Quick Start Manual
Usage Warning:
a) The treatment process should not exceed 20 minutes for every nebulizing, misuse would possibly cause
Pulmonary edema or water poisoning.
b) To prevent superinfection, please pay close attention to the sterilization of nebulizer, ambient air and all
the related equipment, also disinfect animal mouth, nose and throat.
Mechanism of action
Animal respiratory system is an open system, where alveolus is the main air-exchanging place. The object
sprayed out of the nebulizer is mainly the mixture of oxygen and medicine – oxygen-medicine aerosol. This
kind of aerosol could improve oxygen metabolism and anoxic metabolism significantly; increase the oxygen
reserve in organisms. The cavitation effect of compressed air could disperse the liquid medicine into gas
phase and turn the liquid into fog form. The suspended particles will get deep into animal respiratory system
and treat the focus of infection directly. The feature of air compressing nebulization technique is generating
Big Fog Volume Fog with Ultra Small Particle Size (0.5 -5μm); the even fog could go very deep into dendritic
bronchus and the particles could settle down in deep respiratory system, achieves painless cure effect. It is
a good helper to improve curative effect and accelerate recovery process.
Tips:
a) The smaller the particles are, the better curative effect the nebulization could achieve.
b) Normally the fog with bigger particles generated from traditional ultrasonic nebulizer would be easier to
be observed, as the fog seems to be thicker.
c) The international of human medical nebulizer is the particles in fog need to be smaller than 5μm.
Specialized Medical Nebulizer for Animal – Be�er Cura�ve Effect
850EA Air Compressor Nebulizer uses high pressure gas jet knockout the medicine solution into tiny
particles and carry them into deep respiratory system. The size of droplet will affect the settlement site of
medical fog.
Droplet Size
1-5μm
Dendri�c Bronchus, Fine Bronchus, Alveolus
5-20μm
20-40μm
Upper Respiratory Tract mainly, Nose, Throat, etc.
The patented Energy Saving Low Noise Compressor (less than 25 decibels) generates high speed air jet
from filtered clean air (by anti-bacterial cotton), the air stream disperses the medicine solution from liquid
cup into tiny particles and form medicine aerosol. The average size of droplet could reach 2.5μm minus
which could reach deep respiratory system like alveolus and fine bronchus. The medicine is nebulized
evenly, no diluted needed, so there is no residues left almost and could achieve very high medicine
utilization rate. This device is preciously made medical equipment with very low faulty rate and very good
durability, could serve for 1000 -1500 hours as minimum lifespan.
Fig.9.2.1
Fig.9.2.2
Fig.9.2.4
Se�lement Site
Bronchus
Fig.9.2.3
Fig.9.2.5
04
Operation Instructions:
850EA nebulizer is easy to operate, please follow the simple steps as follows.
1. As showed in Figure 9.2.1, open the nebulizer grid, remove the nebulizer cup from the C clip. If the
operator is skilled at this process, he/she could also follow the second step without taking the cup off.
2. Figure 9.2.2, screw off the cup lid and nozzle; pour medicine liquid into the nebulizer cup.
3. Figure 9.2.3, screw on the cup lid and nozzle, and put them back to the left side nebulizer chamber.
4. Figure 9.2.4, Press Nebulize button on control panel to switch on nebulization, the medical fog will be
sprayed through grid and get into the ICU chamber. There are two stages of nebulization treatment timing,
in which the basic time is 10 minutes. Once the nebulization process started, it is 10 minutes countdown
automatically; pressing the "-" button on the nebulize adjustment button will turn it into 20 minutes mode,
pressing the "+" will turn back to 10 minutes mode. Pressing Nebulize button again on the main control
panel when the nebulizing process is running will switch the nebulizer off.
Tips:
a) There is hissing noise when nebulization starts, that's generated from the high speed air flow. The
patented low noise air compressing technology ensures no vibration and no electronic noise when nebulizer
runs. The noise is much lower than normal level of most air compressor nebulizer, which is around 85
decibels normally.
b) When nebulization is ON, the ventilation fans will be turned off to ensure good nebulizing effects, and
humidity and temperature control will be turned off too. The humidity alarm will Not be triggered in nebuliza-
tion mode, which will reset in 3 minutes after nebulization function turned off.
c) The nebulizer panel should be placed at right position as showed in this picture. See Figure 9.2.5
Product Principles and Functions
1. Mechanism of AEOLUS UC-1801 ICU.INCUBATOR
The air circulation system as shown in Figure 5.1.1 is
driven by the circulating air of AEOLUS UC-1801
ICU.INCUBATOR (Pet ICU/Young Pet Incubator). After
being filtered by the purifier, the air from the outside
enters into the ICU, and "dual duct and double
circulation convection heating technology" is adopted to
create an excellent and bacteria-free ICU therapeutic
environment and young pet incubation environment. It
provides clean air and proper temperature and humidity.
2. Main Technical Indicators
1) Power supply voltage: 220V±10%/50Hz±2%
2) Input power: ≤400VA
3) Ambient temperature: 10°C ~ 35°C
4) Temperature control range
Cabinet temperature: 15°C ~ 38°C (can be up to 39°C by special operations)
5) Temperature fluctuation: ≤0.8°C
6) Average temperature of monitoring chamber: ≤1.0°C
7) Cabinet temperature control accuracy: ≤±0.5°C
8) Temperature rise time: 5min ~ 20min
9) Noises in monitoring room: ≤30dB
10) Whole machine earth leakage current: ≤0.5 mA (normal state) ≤1 mA (single fault state)
11) Withstand voltage: 1500V/50Hz, lasted one minute without breakdown and flashover.
12) Ambient conditions: ① Transportation and storage: a. Ambient temperature: -10°C~40°C
② Operating conditions: a. Ambient temperature: 18°C ~ 30°C
b. Relative humidity: ≤95%
b. Relative humidity: 40% ~65%
05