AIR/WATER COOLED HOUSING

The air/water cooled housing shown below allows the sensor to withstand high ambient temperatures.
It is equipped with two 1/8" BSP fittings. Water temperature should be 10°C to 27°C for efficient cool-
ing. Chilled water below 10°C is not recommended. To avoid condensation, the air purge collar should
be used with the water-cooled housing. Water flow rate should not be more than 0.5 to 1.5 litres/min.
1/8 BSP water/air connections
19

AIR PURGE COLLAR

The air purge collar below is used to keep dust, fumes, moisture, and other contaminants away
from the lens. It must be screwed in fully. Air flows into the 1/8" BSP fitting and out of the front
aperture. Air flow should be no more than 5 to 15 litres/min.
Clean or 'instrument' air is recommended.
25
APSW
(for PCAN21)
APSN (for PCAN201)
63
103
1/8 BSP Air connection