N . A S S E I \ 4 B L Y
O Check the Power Source Voltage and Fuse
o Dlsconnect the power cord from the AC input connector. (Make
sure that t rs disconnected.)
. Make sure thatthe line voltage indication plate atthe rear ofthe base
indicates the correct ine vo tage to be used (1OO-120V or
2 2 O - 2 4 O V ) \ F i s . 2 1 ) .
When t indicates Lncorrecl voliage, rerriove the ne voltage
indication plate by unscrewing the two screws using a plus screw
driver.
. Mount the ine voltage ndication p ate after swltch ng lhe input
vo tage change over sw tch to lndlcate correct vo tage.
a Remove the fuse holder at the boitom ofthe base us ng a plus screw
driver (Fig.2 2) and nrake sure that the metal part of the fuse
nd cales
"250V
1 A' for 1 0O 1 20V or
"250V T1
A' tot 224 244V.
When ncorreci fuse type s to be nserted, please contact
your aulhorized Nikon disir bulor.
.Insert ihe fuse into the fuse holder and iightly screw it into the
origina positon using a plus screwdriver.
O Leveling Foot Screw
. Forstable installation of the microscope, man pulatetheadjustment
screw at lhe rear-righi ofthe base (F g. 2 3).