Nikon SMZ1000 Instruction Manual - Page 33

Browse online or download pdf Instruction Manual for Microscope Nikon SMZ1000. Nikon SMZ1000 40 pages. Stereoscopic zoom
Also for Nikon SMZ1000: Instructions For Use Manual (42 pages)

Nikon SMZ1000 Instruction Manual
The optical axis of the 1/4 lambda plate is at
4
the 45° position from the red dot. Brightness
is the maximum when the red dot is
positioned at the front and switches from
bright to dark by each 45° position. Adjust
the plate to the position where the brightness
and contrast are best suited for the
observation.
Insert the filter P-ICI33 NCB14 into the fiber
5
illuminator to improve the color
reproducibility when taking a
photomicrogaph. Refer to the instruction
manual of Fiber Illuminator.
Illuminator
Bright
Dark
Bright
Dark
Bright
V
Red dot
Dark
Bright
Dark
31