Arometrix Extraction Finder User Manual - Page 8

Browse online or download pdf User Manual for Measuring Instruments Arometrix Extraction Finder. Arometrix Extraction Finder 20 pages.
Also for Arometrix Extraction Finder: User Manual (14 pages), Quick Start Manual (2 pages)

Arometrix Extraction Finder User Manual
Design Notes:
The sensor electronics are bathed in an inert nitrogen environment to further insulate them
from any risk of exposure, even though the sensor is intrinsically safe. Solvent will not touch the electronics.
The 30' sensor cable and 10' power cord allows the monitor to be easily placed away from the C1D1 area.
Section 4: Unpacking & Inspecting
After the instrument is received, it should be carefully unpacked and inspected for damage during
shipment and to confirm that all components are present.
Each EXTRACTION FINDER package comes with:
● Extraction Finder Sensor
● Arometrix Monitor
● Sensor Cable, 30'
● Power Supply, 10'
Each EXTRACTION FINDER 2022 and newer package comes with:
● Extraction Finder Sensor
● Arometrix Monitor
● Sensor Cable, 30'
● Power Supply, 10'
● Adapter (optional)
Section 5: Installation & Operation
Warning: Installing a valve or any device that could cause isolation in the line violates NFPA 58. Piping
that can contain liquid LP-Gas and that can be isolated by valving requires a hydrostatic relief valve, as
specified under Section 6.15.
Installation of the Extraction Finder Sensor involves the following steps (also depicted in our Quick Start
Guide):
1. First, mount the Extraction Finder Sensor to the frame of the equipment. An example is shown.
Arometrix, Inc.
arometrix.com/resources
8