Confi guration
Address
Description
Data type #
40517
DATE
UINT32
40519
TAG
CHAR[8]
40523
INPUT1
UINT8
12
Default
Description
0
Confi guration date
(UTC time stamp in seconds starting 1.1.1970)
"V604s"\0
Device text
or
"VB604s"\0
0
00h
Type of measurement Input 1
at 2xmA:
FFh: Measurement is inactive
40h
Wiring variant A
00h:
Voltage measurement [mV]
04h:
Thermocouple internally compensated [K]
60h:
Thermocouple with ext. cold junction thermostat [K]
21h:
Resistance thermometer 2-wire [K]
22h:
Resistance thermometer 3-wire [K]
23h:
Resistance thermometer 4-wire [K]
24h:
Thermocouple with ext. Pt100 on Terminals 1-4 [K]
44h:
Thermocouple with ext. Pt100 on Terminals 2-8 [K]
01h:
Resistance measurement 2-wire [Ω]
02h:
Resistance measurement 3-wire [Ω]
03h:
Resistance measurement 4-wire [Ω]
42h:
Resistance teletransmitter WF
62h:
Resistance teletransmitter WFDIN [Ω]
20h:
Voltage measurement [V]
40h:
Current measurement [mA]
06h:
Sensor earthed: Voltage measurement [mV]
07h:
Sensor earthed: TC internally compensated [K]
66h:
Sensor earthed: TC, ext. cold junction thermostat [K]
27h:
Sensor earthed: TC with ext. Pt100 on Terminals 1-4 [K]
Wiring variant B
10h:
Voltage measurement [mV]
14h:
Thermocouple internally compensated [K]
70h:
Thermocouple with ext. cold junction thermostat [K]
31h:
Resistance thermometer 2-wire [K]
32h:
Resistance thermometer 3-wire [K]
54h:
Thermocouple with ext. Pt100 on Terminals 1-4 [K]
34h:
Thermocouple with ext. Pt100 on Terminals 2-8 [K]
11h:
Resistance measurement 2-wire [Ω]
12h:
Resistance measurement 3-wire [Ω]
52h:
Resistance teletransmitter WF
72h:
Resistance teletransmitter WFDIN [Ω]
16h:
Sensor earthed: Voltage measurement [mV]
17h:
Sensor earthed: TC internally compensated [K]
76h:
Sensor earthed: TC, ext. cold junction thermostat [K]
50h:
2nd current input [mA]
Combination limits are separately shown in a table
on page 16/17.
1
FF
Sensor type Input 1
FFh: Linear
0:
RTD Ptxxx (e.g. Pt100)
1:
RTD Nixxx
2:
Customer-specifi c characteristic curve (only with NLB)
3:
TC Type B
4:
TC Type E
5:
TC Type J
6:
TC Type K
7:
TC Type L
8:
TC Type N
9:
TC Type R
10:
TC Type S
11:
TC Type T
12:
TC Type U
13:
TC Type W5-W26Re
14:
TC Type W3-W25Re
Automatic parameter correction
Operating instructions V604s Be
[Ω]
[Ω]
2
Terminal
3,4
3,4
3,4
1,4
1,3,4
1,2,3,4
1,3,4
3,4,2,8
1,4
1,3,4
1,2,3,4
1,3,4
1,3,4
6,4
5,4
3,4
3,4
3,4
1,3,4
7,8
7,8
7,8
2,8
2,7,8
7,8,1,4
2,7,8
2,8
2,7,8
2,7,8
2,7,8
7,8
7,8
7,8
6,4