Arthur Grillo DA2000 Series Installation And Operation Manual - Page 11
Browse online or download pdf Installation And Operation Manual for Measuring Instruments Arthur Grillo DA2000 Series. Arthur Grillo DA2000 Series 12 pages. Differential pressure manometer
8.
Specification
Manometer
Measurment range:
Display:
Overload protection:
Static pressure:
Pressure connection:
Housing:
Ambient temperature:
Protection class:
Weight:
Tolerance:
Pressure switch
Adjustable switching ranges:
Tolerance on the upper and
lower switching point:
Switching capacity:
Durability:
Contact arrangement:
Analog output
Signal range:
Electrical connection:
Protection class:
0...100 Pa, 0...200 Pa, 0...500 Pa, 0...1000 Pa, 0...2000 Pa,
0...5000 Pa
Scale length 270° = approx. 250 mm
not less then 10-times measurment span
max. 0,2 bar
hose nozzle 8 mm ø
Ultramid/ABS, black, transparent cover
round outside 134 mm ø
-10...50 °C
IP 65
approx. 400 g
± 2 %
20...300 Pa, 30...400 Pa 50...500 Pa, 200...1000 Pa, 500...2500
Pa, 1000...5000 Pa
± 15 %
max. 1,5 A / 250 Vac
mechanical more then 10
normally closed contact - NC
normally open contact - NO
supply - COM
0...10 V, 3-wire-circuit
supply voltage 15...30 Vdc or 24 Vac
4...20 mA, 2-wire-circuit
supply voltage 12...30 Vdc
rectangle plug connector 16 mm or
color coded cable, approx. 50 cm long
IP54 - DA2000, DA2000-A, DA2000-K
IP66 - DA2000-S, DA2000-A-S, DA2000-K-S
Doc.-no: DA2000-01
TECHNICAL SPECIFICATIONS
cycles
6
issue: 02/2016
11