Guidance and Manufacturer's Declaration – Electromagnetic Immunity
Immunity
Test
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
Electrical Fast
Transient/
Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Power
Frequency
(50/60Hz)
Magnetic
Fields
IEC 61000-4-8
12
IEC 60601
Test Level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV differential
mode
±2 kV common
mode
<5% U (>95% dip in
TU) for 0.5 cycle
40% U (60% dip in
TU) for 5 cycles
70% U (30% dip in
TU) for 25 cycles
<5% U (>95% dip in
TU) for 5 sec
3 A/m
Compliance
Level
±8 kV contact
±15 kV air
N/A. Battery
Powered.
N/A. Battery
Powered.
N/A. Battery
Powered.
30 A/m
Electromagnetic
Environment
Guidance
Floors should be
wood, concrete
or ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30%.
Main power
quality should be
that of a typical
commercial
or hospital
environment.
Main power
quality should be
that of a typical
commercial
or hospital
environment.
Main power
quality should be
that of a typical
commercial
or hospital
environment.
Power frequency
magnetic fields
should be at levels
characteristic of
a typical location
in a typical