CUSTOM ULTRASONICS 20000 Operator's Manual - Page 7

Browse online or download pdf Operator's Manual for Storage CUSTOM ULTRASONICS 20000. CUSTOM ULTRASONICS 20000 7 pages. Endoscope storage cabinet
Also for CUSTOM ULTRASONICS 20000: Operator's Manual (11 pages)

CUSTOM ULTRASONICS, INC.
6.
Operator Use or Misuse related to potential hazards may become humid and failure to:
a. Inspect the desiccant bags and humidity indicator cards daily
i. If not checked and the adsorbent material in the desiccant bags become
saturated, the cabinet may become humid and not facilitate drying of the
endoscope
b. Properly connect the male bulkhead adapter fittings to the cabinet's female bulkhead
purge manifold connector.
i. The storage cabinet connector must also be attached properly to the
endoscope.
ii. If either end is not connected, the internal channels of the endoscope(s) will
not have the proper air flow to facilitate drying.
7. Daily Cleaning:
a. The stainless steel drip plate [located on the inside of the storage cabinet base] must
be cleaned and sanitized daily using a damp cloth and a standard hospital grade, non-
abrasive, surface disinfectant/cleaner.
b. Check inside of cabinet for rust or peeling paint. If noted, immediately repair the
surface and repaint. Contact
8. Daily Observations:
a. Once a day, press the momentary override switch located on the back of the silver
bar containing the ON/OFF switch and timer.
i. When activating this switch, check to ensure that the air pump and vent fan
are turned on and are operational.
b. Once a day, listen for the operation of the air pump to confirm that it is coming on
by the automatic timer.
i. The pump should run for 10 minutes or the pre-set time noted on the dial on
the front of the cabinet.
ii. The pump is factory set at 10 minutes, but can be changed according to
hospital/facility requirements.