DLAB Laboratory Instruments DM0412S User Manual - Page 15

Browse online or download pdf User Manual for Laboratory Equipment DLAB Laboratory Instruments DM0412S. DLAB Laboratory Instruments DM0412S 18 pages. Clinical centrifuge

DLAB Laboratory Instruments DM0412S User Manual
to the manufacturer's instructions. Use detergent with its pH between 7 and 9.
2)
Autoclaving PA、PC and PP tubes
2
PA begins softening at about 120℃, PC and PP at about 130℃. Autoclave PA tubes at 115℃(0.7kg/cm
2
for 30 minutes, PC and PP tubes at 121℃ (0.1kg/cm
) for 20 minutes. If a certain temperature is exceeded,
the tubes may be deformed.
When use a sterilizing chamber, please operate as follows:
(1) Place tubes in vertical position, mouths upward. If tubes are placed sideways, they may deform into an
oval shape due to gravity.
(2) Remove locking nut and lid to prevent from deformation or rupture.
(3) Wait until the sterilizing chamber cools down to the room temperature before removing tubes.
4) Conditions and life expectancy of tubes
The life expectancy of plastic tubes depends on the characteristics of samples, speed of the rotor used,
temperature applied and so on. When the plastic tubes are used for ordinary aqueous samples (pH between
5.0 and 9.0), their life expectancies are defined as follows.
Be operated at the maximum speed:
High quality tubes (PA、PC、PP): 30-50 operations
Ordinary tubes(PA、PC、PP): around 10 operations(Using in low speed can extend the tube life).
Life expectancy of tubes also depends on the pretreatment conditions such as cleaning and sterilization,
lifetime can be cut down.
Notice: Do not use damaged or cracked tubes.

11. Returning and Disposal

11.1 Returning Devices

Before returning the device, a transport securing device has to be installed.
If the device or its accessories are returned back, in order to provide protection for people, the environment
and materials, it has to be decontaminated and cleaned before being
shipped.

11.2 Disposal

Before disposal, the device must be decontaminated and cleaned to protect people, the environment and
property. When you are disposing of the device, the respective statutory rules must be observed.
Pursuant to guideline 2002/96/EC (WEEE), all devices supplied after August 13,2005 may not be disposed